EUROLAB laboratory provides testing and compliance services within the scope of EN 61010-1 standard. This standard specifies general safety rules for electrical equipment types and accessories, such as electrical test and measurement equipment, electrical industrial process control equipment, electrical laboratory equipment.
This standard does not apply to equipment covered by the following standards:
The purpose of the rules specified in the EN 61010 standard is to reduce the hazards to the operator and his environment to a tolerable level.
Safety Test (LVD Test)
Electrical safety tests (LVD testing) are tests to verify the compliance of any commercial electrical product with relevant standards. Various governments and organizations have developed requirements for electrical products sold worldwide. In most markets, a product must meet safety standards (LVD Test).
To verify compliance with these standards (LVD Test), the following product must pass the safety tests;
Dielectric Strength Test
It is applied between the power circuit of the test sample and the protective earth circuit. The test voltage must have a rated frequency of 50 Hz or 60 Hz. The largest test voltage should be applied to the insulations to check the effectiveness of the insulations.
Insulation resistance test
The insulation resistance measured after 500 V direct current (DC) applied between the power circuit conductors and the ground circuit of the test sample must be greater than 100 MΩ.
Soil Continuity Test
At ground continuity, the reliability of the grounding system of the test sample is checked. The resistance of each protective earth circuit between the protective earth terminal and the areas where the sample is expected to be earthed should be measured with a suitable test equipment. The test result is evaluated as positive or negative according to the values specified in the standard.
Leakage Current Test
Leakage current testing is the measurement of the current flowing between the accessible surfaces of the test specimen and the ground with a suitable test device under normal, abnormal and fault conditions. As a result of the test, controls are made within the limits given by the standard.
EUROLAB, with its more than 25 years of experience, state-of-the-art accredited laboratories and expert team, helps you get precise and fast results.
To get an appointment, to get more detailed information or to request an evaluation, you can ask us to fill in our form and reach you.